Patient information: uses, side effects, how to take - Ultra-Technekow V4
Directions for Use of the Technetium Tc 99m Generator
NOTE 1: Immediately upon delivery, the generator should be placed within a minimum of one-inch of lead shielding in such a manner
so as to minimize radiation exposure to attending personnel.
NOTE 2: Wear waterproof gloves during the elution procedure and during subsequent reconstitution of kits with the eluate.
NOTE 3: Use a shielded syringe to withdraw patient dose or to transfer Sodium Pertechnetate Tc 99m into mixing vials during kit
reconstitution.
NOTE 4: The needles in the generator are sterile beneath their covers, and the generator has been cleaned underneath the top cover.
Additional disinfection of these areas with agents containing alcohol may unfavorably influence the Tc-99m yield.
Eluting the generator every 24 hours will provide optimal amounts of Sodium Pertechnetate Tc 99m. However, the generator may be eluted whenever sufficient amounts of technetium Tc-99m have accumulated within the column.
For Example
Time After First Elution (hrs.) | Approximate Yield (% of First Elution) |
| 1 | 10 |
| 2 | 19 |
| 3 | 27 |
| 4 | 35 |
| 5 | 41 |
| 6 | 47 |
Elution
- 1. Lift the generator by its handle and place it inside the auxiliary shield. Move the handle so that it is not covering the generator top by pushing it off to the side in between the generator and the auxiliary shield.
- 2. Remove and store the elution hood cover. Place the auxiliary shield top onto the top of the generator and align it with the elution hood.
- 3. Remove and store the tip cap plugs from the needles.
- 4. Remove the flip-top cap of the eluant vial; disinfect the stopper with a bacteriocide such as 70% isopropyl alcohol, allowing the stopper to dry before use. Invert the eluant vial and place stopper first into the saline vial alignment insert. Place the saline vial alignment insert and vial into the saline port of the auxiliary shield top and firmly push down the eluant vial until the stopper is punctured and seated at the base of the eluant needles.
- 5. Place the saline shield on top of the auxiliary shield top to cover the eluant vial.
- 6. Remove the flip-top cap of an evacuated vial; disinfect the stopper, allowing the stopper to dry before use. Place the evacuated vial into the elution tool.
- 7. Position the shielded evacuated vial by carefully lowering the elution tool into place on the elution needle. Piercing the septum of the evacuated vial with the elution needle will begin the elution process.
- 8. Wait until the evacuated vial has completely filled itself. This may take a few minutes. Never interrupt the elution by lifting the elution tool!
- 9. NOTE: Do not use generator eluate if its appearance is discolored, and discontinue use of the generator.
- 10. Remove the flip-top cap of the Technestat vial; disinfect the stopper, allowing the stopper to dry before use. Secure the Technestat vial into the Technestat vial holder.
- 11. Carefully remove the elution tool and replace with the shielded Technestat vial.
- 12. Determine the technetium Tc-99m concentration and molybdenum Mo-99 content for dispensing purposes. The generator eluate may be assayed using an appropriate detection system. The manufacturer’s instructions for operation of the instrument/equipment should be followed for measurement of Technetium Tc-99m and Molybdenum Mo-99 activity. NOTE: Molybdenum Mo-99 Breakthrough Limit- The acceptable limit is 0.15 kilobecquerel molybdenum Mo-99 per megabecquerel technetium Tc-99m (0.15 microcurie Mo-99 per millicurie Tc-99m) per administered dose in the Injection, at the time of administration (see USP, Sodium Pertechnetate Tc 99m Injection).
- 13. Determine the aluminum ion concentration of the eluate. NOTE: Aluminum Ion Breakthrough Limit- The acceptable limit is not more than 10 micrograms per milliliter of eluate (see USP, Sodium Pertechnetate Tc 99m Injection).
Subsequent Elutions
- 1. Remove the saline shield and then remove saline vial alignment insert from the saline port to remove the eluant vial from the eluant needles. Remove the vial from the saline vial alignment insert and reuse the saline vial alignment insert for subsequent elutions.
- 2. Remove the shielded Technestat vial by carefully lifting the Technestat vial shield from the elution needle.
- 3. Repeat steps 4 through 12 of the Elution procedure above.
Vacuum Loss
If the vacuum in the collecting vial is lost, do not attempt to re-evacuate the vial, but discard and use a new collecting vial.
Expired Generator Disposal
- 1. Following the life of the generator, remove and dispose of the used Technestat vial and the eluant vial.
- 2. Cover the elution and eluant needles with the stored tip cap plugs.
- 3. Place the stored elution hood cover onto the top of the generator.
- 4. The intact generator assembly should be either returned to Curium US LLC or disposed of in accordance with applicable regulations.
This generator may be received, used and administered only by authorized persons in designated clinical settings. Its receipt, storage, use, transfer and disposal are subject to the regulations and/or appropriate licenses of local competent official organizations.
©2022 Curium US LLC. Ultra-Technekow TM, Technestat TM, Curium™, and the Curium logo are trademarks of a Curium company.
Manufactured by:
Curium US LLC
Maryland Heights, MO 63043 USA
Made in USA
A901IS
Revised 5/2022
STERILE
CURIUM™